Agreement between optical coherence and Scheimpflug tomography: Vault measurements and reproducibility after
I Almorín-Fernández-Vigo1, I Sánchez-Guillén2, J I Fernández-Vigo3
1Centro Internacional de Oftalmología Avanzada, Badajoz, Spain.
Purpose:
To evaluate the agreement between Scheimpflug tomography (Pentacam, Oculus) and anterior segment optical coherence tomography (AS-OCT, RTVue 100, Optovue) as well as the reproducibility of each technique in assessing the lens-ICL distance (vault) after implantable collamer lens (ICL) implantation.
Methods:
The vault was measured manually with Scheimpflug tomography and AS-OCT. Intraclass correlation coefficients (ICC) and Bland-Altman plots were used to determine the reproducibility of measurements and the agreement between them. Multivariate regression analysis was performed to identify predictors of differences in vault measurements between devices.
Results:
80 eyes of 46 ICL patients were analyzed. Mean patient age was 33.8±7.4 years (range, 21 to 51), and 27 (59%) were women. The preoperative spherical equivalent refraction ranged from -26 to 8.5 diopters (D). The mean vault measured by AS-OCT was 558.8±240.2μm (range, 162 - 1220) vs. 430.1±215.3μm (range, 0 to 1070) by Pentacam. AS-OCT yielded vault values, on average, 128.1±64.6μm higher than the Pentacam (range, -22 to 293μm). The ICC values for consistency and absolute agreement were 0.960 and 0.928, respectively. Reproducibility of vault measurements was excellent for both devices (ICC≥0.946). Four parameters-vault, pupil diameter (PD), PD differences between devices, and preoperative spherical equivalent refraction-were significant predictors of differences in vault measurements (adjusted-R2=0.412; P<0.001).
Conclusions:
Agreement between AS-OCT and Pentacam for vault measurement was good. Vaults were higher when measured by AS-OCT. These differences are clinically relevant, and therefore these devices cannot be used interchangeably to measure vault.
More Related Videos
12:22Multimodal Volumetric Retinal Imaging by Oblique Scanning Laser Ophthalmoscopy oSLO and Optical Coherence Tomography OCT
Published on: August 4, 2018
12:54Simultaneous Brightfield, Fluorescence, and Optical Coherence Tomographic Imaging of Contracting Cardiac Trabeculae Ex Vivo
Published on: October 2, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography