Online Solid-Phase Extraction-Inductively Coupled Plasma-Quadrupole Mass Spectrometry with Oxygen Dynamic Reaction
Makoto Matsueda1,2, Kayo Yanagisawa1, Kazuma Koarai2
1Faculty of Symbiotic Systems Science, Cluster of Science and Technology, Fukushima University, 1 Kanayagawa, Fukushima 960-1296, Japan.
Abstract:
Quantification of pg/L levels (i.e., 0.6 mBq/L) of radioactive technetium-99 (99Tc) was achieved within 15 min in the presence of isobaric and polyatomic interference sources such as ruthenium-99 (99Ru) and molybdenum hydride (98Mo1H) at 3-11 orders of magnitude higher concentrations. Online solid-phase extraction-inductively coupled plasma-quadrupole mass spectrometry (ICP-QMS) with oxygen (O2) dynamic reaction cell (online SPE-ICP-MS-DRC) was shown to be a thorough automatic analytical system, circumventing the need for human handling. At three stepwise separations (SPE-DRC-Q mass filters), we showed that interference materials allowed the coexistence of abundance ratios of 1.5 × 10-13 and 1.1 × 10-5 for 99Tc/Mo and 99Tc/Ru, respectively. A classical mathematical correction using the natural isotope ratio of 99Ru/102Ru was used to calculate the residues of 99Ru. Using this optimized system, a detection limit (DL; 3σ) of 99Tc was 9.3 pg/L (= 5.9 mBq/L) for a 50 mL injection and sequential measurements were undertaken at a cycle of 24 min/sample. For the measurement of a lower concentration of 99Tc, an AG1-X8 anion-exchange column was used to study 20 L of seawater. Its DL was approximately 1000 times greater than that of previous methods (70.0 fg/L). Thus, this method withstands coexistences of 5.8 × 10-18 and 3.5 × 10-9 for 99Tc/Mo and 99Tc/Ru, respectively. Spike and recovery tests were conducted for environmental samples; the resulting values showed good agreement with the spike applied.
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
Mass Analyzers: Common Types
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences