Applying X-ray Imaging Crystal Spectroscopy for Use as a High Temperature Plasma Diagnostic
Published on: August 25, 2016
A von Hamos spectrometer based on highly annealed pyrolytic graphite crystal in tender x-ray domain
Iyas Ismail1, Loïc Journel1, Régis Vacheresse1
1Sorbonne Université, CNRS, UMR 7614, Laboratoire de Chimie Physique-Matière et Rayonnement, 4 Place Jussieu, 75252 Paris, France.
Abstract:
We have built an x-ray spectrometer in a von Hamos configuration based on a highly annealed pyrolytic graphite crystal. The spectrometer is designed to measure x-ray emission in the range of 2-5 keV. A spectral resolution E/ΔE of 4000 was achieved by recording the elastic peak of photons issued from the GALAXIES beamline at the SOLEIL synchrotron radiation facility.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...