Biomechanical comparative evaluation of percutaneous fixations with vertebral expansion for vertebral compression
Le Gallo Lucas1, V P C Lima Lucas1,2, Persohn Sylvain1
1Institut de Biomecanique Humaine Georges Charpak, Paris, France.
Abstract:
This study uses in vitro experiments and validated finite element models (FEM) to analyze the effect of posterior fixation, alone or associated with expandable device (ED) and/or cement. 3-dimensional FEMs of intact, fractured and instrumented spine were built and compared with experimental load-displacement curves. FEM ranges of motion were within the experimental corridors. Stresses appeared sensitive to both implant configuration and fracture severity with a stress reduction up to 84%. The FEM highlighted that for a same instrumental strategy, different biomechanical performances were observed according to fracture severity. When bone continuity is altered, both ED and cement may be needed.