Short report: Performance evaluation of the Idylla™ KRAS and EGFR mutation tests on paraffin-embedded cytological
Saskia Offerman1, Clemens F Prinsen1,2, Ageeth Knol1
1Isala Pathology, Dr. Van Heesweg 2, 8025 AB Zwolle, Postbus 10400, 8000 GK, Zwolle, The Netherlands.
Background:
Quick and reliable testing of EGFR and KRAS is needed in non-small cell lung cancer (NSCLC) to ensure optimal decision-making for targeted therapy. The Idylla™ platform was designed for Formalin-Fixed Paraffin-Embedded (FFPE) tissue sections but recently several studies were published that evaluated its potential for cytological specimens. This study aimed to validate the Idylla™ platform for the detection of EGFR/KRAS mutations in cytological NSCLC samples prepared as cytoblocks using AGAR and paraffin embedding.
Material And Methods:
The KRAS Idylla™ test were performed on 11 specimens with a known KRAS mutation. The EGFR Idylla™ test was performed on 18 specimens with a known primary EGFR mutation and 7 specimens with a primary EGFR-EGFR T790M resistance mutation combination.
Results:
Concordant KRAS and primary EGFR mutations were detected for both KRAS and primary EGFR mutations. Samples with a total CQ value of < 26 could be considered negative. Samples with a total CQ value of > 26 could not be assessed (probability of false-negative). In specimens with a primary EGFR-EGFR T790M resistance mutation combination, 5/7 cases were not concordant.
Conclusion:
Our results confirm the conclusion of recent reports that the Idylla™EGFR assay is not suitable in a resistance to EGFR TKI setting, also not in our cytological NSCLC samples prepared as cytoblocks using AGAR and paraffin embedding. KRAS and primary EGFR mutations were detected using the Idylla™ assays in virtually all cytological NSCLC samples. This analysis was rapid and time-saving compared to other mutation detection assays and may be useful if the amount of material is insufficient to perform a full set of molecular tests.