The biotin interference within interference suppressed immunoassays
Payam Kabiri1, Ralf Weiskirchen1, Josef van Helden1,2
1Institute for Molecular Pathobiochemistry, Experimental Gene Therapy and Clinical Chemistry (IFMPEGKC), University Hospital RWTH, Aachen, Germany.
Biotin supplements can interfere with laboratory tests, leading to inaccurate results. This study found that biotin interference in immunoassays often causes falsely elevated, not lowered, test results, impacting medical decisions.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Biochemistry
Background:
- Biotin supplementation is increasingly common, raising concerns about its interference with laboratory immunoassays.
- Immunoassay reliability is critical, as up to 70% of medical decisions depend on laboratory findings.
Purpose of the Study:
- To evaluate the reliability of immunoassays in the presence of biotin interference.
- To investigate the impact of biotin supplementation on laboratory test results.
Main Methods:
- Examined 59 interference-suppressed immunoassays from Roche Diagnostics for biotin interference sensitivity.
- Investigated biotin pharmacokinetics in volunteers supplemented with biotin.
Main Results:
- Biotin interference in laboratory findings is probable.
- Falsely elevated test results are more frequent than falsely low results in sandwich immunoassays, contrary to existing research.
Conclusions:
- Interference-suppressed immunoassays are susceptible to biotin interference.
- Laboratory professionals, medical staff, and patients must be aware of potential biotin interference.
- Roche Diagnostics should review interference suppression and test indications.