Related Experiment Video
Updated: Aug 11, 2026

Determination of Microbial Extracellular Enzyme Activity in Waters, Soils, and Sediments using High Throughput Microplate Assays
Published on: October 1, 2013
Use of a 96-well microplate reader for measuring routine enzyme activities
M B Ashour1, S J Gee, B D Hammock
1Department of Entomology, University of California, Davis 95616.
Abstract:
A method is described for the routine determination of the rate of colorimetric enzyme reactions using a 96-well microtiter plate reader commonly used in immunoassay. This approach is illustrated by monitoring esterase activity using three common products: release of thiol, release of ethanol, and release of p-nitrophenylate ion. Examples include monitorings of the rate of hydrolysis of acetylthiocholine iodide by eel acetylcholinesterase and the rate of hydrolysis of malathion and nonconventional esters such as O-methyl, O-ethyl, and O-isobutyl carbonates of p-nitrophenol by commercial porcine liver carboxylesterase. Data obtained from the plate reader were compared to those obtained, under similar conditions, in a conventional spectrophotometer. Absorbance measurements made in both machines on the same solution, as well as absorbance changes measured over time, were similar. The use of the 96-well plate format tremendously increased the number of enzyme assays carried out per person and the interface with a personal computer allowed rapid manipulation of the absorbance values to calculate the desired rate data. This approach should be generally applicable to many routine colorimetric assays in the research laboratory.