Related Experiment Video
Updated: Oct 25, 2025

Intravenous Microinjections of Zebrafish Larvae to Study Acute Kidney Injury
Published on: August 4, 2010
Elimination of three doses of gentamicin over three consecutive days using a polyacrylonitrile-derived filter: An in
Frédéric J Baud1,2,3, Vanessa Seif4, Pascal Houzé5
1Adult Intensive Care Unit, Department of Anesthesiology - SAMU de Paris, Assistance Publique - Hôpitaux de Paris, University Hospital Necker, Paris, France.
Introduction:
Adsorption of gentamicin in a polyacrylonitrile filter was previously evidenced in a session lasting 6 h using the NeckEpur model. We extended the study over three consecutive days to mimic the 72-h life span of a filter.
Methods:
Prismaflex® monitor and ST150® filter were used in the continuous diafiltration (CDF) mode at a 2.5 L/h flowrate. The daily session started with a 6-h session of CDF. Thereafter, the 5-L central compartment was changed using a bag free of gentamicin to assess gentamicin release over the following 18 h. Experiments were repeated on Day 2 and stopped at the end of the 6-h session of CDF on Day 3. The experiment was performed in duplicate.
Results:
At a 2.5 L/h diafiltration flowrate, the mean daily clearances of gentamicin were 5.5, 4.0, and 3.3 L/h, respectively. The mean diafiltration and adsorption ratios in the daily elimination of gentamicin were 32/68%, 58/42%, and 88/12%, respectively. During days 1 and 2, the mean amount of gentamicin released from the ST150® filter were 14 and 34 mg, respectively.
Conclusion:
The pharmacokinetics of gentamicin over 3 days is strongly altered by adsorption in the same filter with a progressive decrease of elimination by adsorption, suggesting saturation of the filter. One limitation of our study results from the mode of administration using a bolus dose instead of an infusion over 30 min. Adsorption adds a clearance to those of diafiltration. The time-dependency of gentamicin clearance precludes using a constant dosage regimen over the filter's life span.
More Related Videos
12:11A Modified EPA Method 1623 that Uses Tangential Flow Hollow-fiber Ultrafiltration and Heat Dissociation Steps to Detect Waterborne Cryptosporidium and Giardia spp.
Published on: July 9, 2012
06:18A Novel Method to Determine the Longitudinal Antibacterial Activity of Drug-Eluting Materials
Published on: March 3, 2023
Related Concept Videos
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Estimation of k and VD of Aminoglycosides
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy