Comparison of skin dose calculated by the dose tracking system (DTS) with a beam angular correction factor and that
Sheng-Hsuan Sun1, Stephen Rudin1, Daniel R Bednarek1
1The State University of New York at Buffalo, Jacobs School of Medicine and Biomedical Sciences, Canon Stroke and Vascular Research Center, 875 Ellicott St., Buffalo, NY 14203.
Abstract:
Skin dose is dependent on the incident beam angle and corrections are needed for accurate estimation of the risk of deterministic effects of the skin. Angular-correction factors (ACF) were calculated and incorporated into our skin-dose-tracking system (DTS) and the results compared to Monte-Carlo simulations for a neuro-interventional procedure. To obtain the ACF's, EGSnrc Monte-Carlo (MC) software was used to calculate the dose averaged over 0.5, 1, 2, 3, 4 and 5 mm depth into the entrance surface of a water phantom at the center of the field as a function of incident beam to skin angle from 90-10 degrees for beam field sizes from 5-15 cm and for beam energies from 60-120 kVp. These values were normalized to the incident primary dose to obtain the ACF. The angle of incidence at each mesh vertex in the beam on the surface of the DTS patient graphic was calculated as the complement of the angle between the normal vector and the vector of the intersecting ray from the tube focal spot; skin dose at that vertex was calculated using the corresponding ACF. The skin-dose values with angular correction were compared to those calculated using MC with a matching voxelized phantom. The results show the ACF decreases with decreasing incident angle and skin thickness, and increases with increasing field size and kVp. Good agreement was obtained between the skin dose calculated by the angular-corrected DTS and MC, while use of the ACF allows the real-time performance of the DTS to be maintained.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Biological Effects of Radiation