Hidden Markov model-based heartbeat detector using electrocardiogram and arterial pressure signals
Miguel Altuve1,2, Nelson F Monroy3
1Valencian International University, Valencia, Spain.
Insights
This study introduces a novel heartbeat detection method using hidden Markov models (HMMs) to combine electrocardiogram (ECG), arterial blood pressure (ABP), and pulmonary arterial pressure (PAP) signals, achieving over 99% accuracy.
Area of Science:
- Biomedical Engineering
- Signal Processing
- Cardiovascular Monitoring
Background:
- Electrocardiogram (ECG) alone has limitations for reliable heartbeat detection due to noise and data loss.
- Combining multiple physiological signals offers a more robust approach to accurately detect heartbeats.
Purpose of the Study:
- To develop and evaluate a novel heartbeat detection system integrating information from ECG, arterial blood pressure (ABP), and pulmonary arterial pressure (PAP) signals.
- To leverage hidden Markov models (HMMs) for joint signal processing to enhance heartbeat detection accuracy.
Main Methods:
- Physiological signals (ECG, ABP, PAP) were preprocessed.
- A sliding window approach extracted observation sequences for analysis.
- Two trained hidden Markov models (HMMs) were employed to calculate log-likelihoods for detection based on an adaptive threshold.
Main Results:
- The HMM-based detector achieved high performance across multiple databases (MIT-BIH Arrhythmia, PhysioNet Challenge 2014, MGH/MF Waveform).
- Sensitivity and positive predictivity exceeded 99% on two databases and 95% on another, using ECG and ABP signals.
- Detection performance was found to be comparable to existing state-of-the-art methods.
Conclusions:
- The proposed HMM-based approach effectively integrates multi-signal physiological data for accurate heartbeat detection.
- This method offers a reliable and robust alternative to traditional ECG-only detection, especially in the presence of signal noise or missing data.
Abstract:
The automatic detection of a heartbeat is commonly performed by detecting the QRS complex in the electrocardiogram (ECG), however, various noise sources and missing data can jeopardize the reliability of the ECG. Therefore, there is a growing interest in combining the information from many physiological signals to accurately detect heartbeats. To this end, hidden Markov models (HMMs) are used in this work to jointly exploit the information from ECG, arterial blood pressure (ABP) and pulmonary arterial pressure (PAP) signals in order to conceive a heartbeat detector. After preprocessing the physiological signals, a sliding window is used to extract an observation sequence to be passed through two HMMs (previously trained on a training dataset) in order to obtain the log-likelihoods of observation and signals a detection if the difference of log-likelihoods exceeds an adaptive threshold. Several HMM-based heartbeat detectors were conceived to exploit the information from the ECG, ABP and PAP signals from the MIT-BIH Arrhythmia, PhysioNet Computing in Cardiology Challenge 2014, and MGH/MF Waveform databases. A grid search methodology was used to optimize the duration of the observation sequence and a multiplicative factor to form the adaptive threshold. Using the optimal parameters found on a training database through 10-fold cross-validation, sensitivity and positive predictivity above 99% were obtained on the MIT-BIH Arrhythmia and PhysioNet Computing in Cardiology Challenge 2014 databases, while they are above 95% in the MGH/MF waveform database using ECG and ABP signals. Our detector approach showed detection performances comparable with the literature in the three databases.
Supplementary Information:
The online version contains supplementary material available at 10.1007/s13534-021-00192-x.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Holter Monitor: 24-Hour Monitoring
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: