Detection of Human Leukocyte Antigen Biomarkers in Breast Cancer Utilizing Label-free Biosensor Technology
Published on: March 24, 2015
[Two Serum Free Light Chain Detection Systems in the Diagnosis of Multiple Myeloma]
Guo-Qing Zhu1, Xue Fu1, Yan-Song Ren1
1State Key Laboratory of Experimental Hematology, National Clinical Research Center for Blood Diseases, Institute of Hematology & Blood Diseases Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Tianjin 300020,China.
Objective:
To investigate the comparability of the Freelite, Binding Site, Beckman and N Latex FLC, Siemens in the detection of serum free light chain (sFLC) .
Methods:
Fifty newly diagnosed multiple myeloma (MM) patients in Tianjin Institute of Blood Research from November 2019 to February 2020 were enrolled. The two systems (Freelite, Binding Site, Beckman and N Latex FLC, Siemens) were used to detect the sFLC of the samples. Outlier detection was performed by ESD method, methodological comparison and deviation assessment were performed by Passing-Bablok regression and Bland-Altman regression.
Results:
Both the systems could quantitatively analyze free kappa light chain serum samples and free lambda light chain samples. Freelite, Binding Site, Beckman and N Latex FLC, Siemens free light chain test showed FLC-κ:36.5 (6.5, 194), 40.5 (6.94, 288), FLC-λ: 30.1 (4.3, 170.5), 35.1 (2.28, 526), rFLC (FLC-κ/ FLC-λ) : 0.82 (0.05, 43.25), 1.03 (0.03, 32.04), dFLC (|FLC-κ- FLC-λ|) : -5.8 (-161.97, 183.7), 1.1 (-505.1, 279.01), which existed no outliers. There were systematic differences, and the deviation level was not within the clinically acceptable range.
Conclusion:
Both the systems can meet the needs of clinical diagnosis and treatment, but there is a significant deviation between the two systems, the results are not comparable, and should be analyzed separately. In particular, the same system should be selected for monitoring the prognosis of MM.