Automatic versus manual oxygen titration using a novel nasal high-flow device in medical inpatients with an acute
James Harper1,2,3, Nethmi Kearns4,3, Grace Bird4,2,3
1Medical Research Institute of New Zealand, Wellington, New Zealand james.harper@mrinz.ac.nz.
Background:
Guideline recommendations state oxygen should be administered to acutely unwell patients to achieve a target oxygen saturation (SpO2) range. The current practice of manual oxygen titration frequently results in SpO2 outside of a prescribed range. The aim of this study was to assess the efficacy of automatic oxygen titration using a closed-loop feedback system to achieve SpO2 within a prescribed target range METHODS: An open-label randomised parallel group trial was undertaken comparing automatic oxygen titration using a novel nasal high-flow device to manual oxygen titration using nasal high flow. Medical inpatients requiring oxygen therapy in Wellington Regional Hospital, New Zealand with a prescribed target SpO2 range of 88%-92% or 92%-96% were recruited and randomised equally between the interventions for a period of 24 hours. The primary outcome was the proportion of time spent with SpO2 within the prescribed range.
Results:
20 patients were included in the analysis. Automatic oxygen titration resulted in a median (IQR) 96.2% (95.2-97.8) of time within the target range compared with 71% (59.4-88.3) with manual titration; difference (95% CI) 24.2% (7.9% to 35%), p<0.001. There was a reduction in the time spent with SpO2 ≥2% above and ≥2% below range in the automatic titration group, although the point estimate for the differences were small; -1% (-8.2% to -0.04%), p=0.017 and -2.4% (-11.5% to 0.3%), p=0.05 respectively.
Conclusions:
Nasal high-flow with automatic oxygen titration resulted in a greater proportion of time spent with SpO2 in target range compared with manual titration.
Trial Registration:
The trial was registered with the Australian and New Zealand Clinical Trials Registry (ACTRN12619000901101).
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...