Elimination of CO2 insufflation-induced hypercapnia in open heart surgery using an additional venous reservoir
Jesper Nyman1,2, Manne Holm1,2, Thomas Fux1,2
1Division of Perioperative Medicine and Intensive Care, Section Cardiothoracic Surgery and Anesthesiology, Karolinska University Hospital, Stockholm, Sweden.
Objectives:
Carbon dioxide (CO2) gas insufflation is used for continuous de-airing during open heart surgery. The aim was to evaluate if an additional separate venous reservoir eliminates CO2 insufflation-induced hypercapnia and keeps sweep gas flow of the oxygenator constant.
Methods:
A separate reservoir was used during cardiopulmonary bypass in addition to a standard venous reservoir. The additional reservoir received drained blood and CO2 gas continuously via a suction drain (1 l/min) and handheld suction devices from the surgical wound. CO2 gas was insufflated via a gas diffuser in the open wound at 10 l/min. In a cross-over design for each patient, gas and blood were either continuously drained from the additional to the standard venous reservoir or not. CO2 pressure in arterial blood (PaCO2) was measured after adjustment of sweep gas flow as necessary and after steady state of PaCO2 was observed. Mean values for each setup (median 4 times) for each patient were analysed with Wilcoxon rank-sum test.
Results:
Ten adult patients undergoing open aortic valve replacement were included. Median PaCO2 did not differ between setups (5.41; 5.29-5.57, interquartile range vs 5.41; 5.24-5.58, P = 0.92), whereas sweep gas flow (l/min) was lower (2.58; 2.50-3.16 vs 4.42; 4.0-5.40, P = 0.002) when CO2 gas was not drained from the additional to the standard reservoir.
Conclusions:
An additional venous reservoir for the evacuation of blood from the open surgical wound eliminates CO2 insufflation-induced hypercapnia in open heart surgery keeping PaCO2 and sweep gas flow constant. This prevents possible CO2-induced hyperperfusion of the brain and decreases the risk of cerebral particulate embolization during CO2 insufflation for de-airing in open heart surgery.
Clinical Trial Registration:
NCT04202575.
Irb Approval Dat And Number:
2018-07-13 and 2018/1091-31.
More Related Videos
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Acute Respiratory Failure-III
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...