Intraoperative detection of parathyroid glands by autofluorescence identification using image-based system: report of
Maksym Gorobeiko1,2, Andrii Dinets3,4
1Department of Surgery, Taras Shevchenko National University of Kyiv, Demiїvska 13, Ukraine, Kyiv, 03039, Ukraine.
Background:
A detection of parathyroid glands by the evaluation of their autofluorescence in the near-infrared spectrum is considered as a promising tool in addition to their visual verification. The aim of this study was to evaluate the role of near-infrared autofluorescence application by using two different image-based systems for the identification of parathyroid glands during surgery of thyroid and parathyroid benign and malignant lesions.
Methods:
Evaluation of near-infrared autofluorescence was performed in 15 patients by using two different image-based systems equipped with a near-infrared laser camera. Intravenous injection of fluorophore indocyanine green was used for the enhancement of near-infrared autofluorescence signal.
Results:
Normal parathyroid glands were identified and mobilized after visual inspection in 12 (80%) patients, which was confirmed by near-infrared autofluorescence evaluation. Confident recognition of parathyroid glands by near-infrared autofluorescence signal and their subsequent distinction from lymph nodes was achieved in two (13%) patients with prior surgery for papillary thyroid carcinoma. In one (7%) case, parathyroid gland was identified as fragments of tissue within the postoperative scarring area by near-infrared autofluorescence evaluation, but not by visual inspection. A less intensive near-infrared autofluorescence signal was detected in the parathyroid gland owing to unintentional excision in one (7%) case. Better signal intensity from parathyroid glands was noticed after changes of the near-infrared camera in Fluobeam 800 image-based system in position to an angle of approximately 45-65° in relation to area of interest in all cases as compared with holding straight on the parathyroid gland. Fluobeam LX demonstrated a good near-infrared autofluorescence signal without any specific changes in the camera angle. Thyroid carcinoma demonstrated low-intensity signal in the case of invasion to thyroid capsule. No fluorescent signal was identified from metastatic, or from normal, lymph nodes.
Conclusions:
The application of near-infrared autofluorescence imaging is considered as a useful, but additional, tool for the visual assessment of parathyroid gland in the case of primary neck exploration. The utility of near-infrared autofluorescence imaging for parathyroid detection is increased in the case of repeated surgical intervention owing to increased risk of unintentional parathyroid removal as well as for discrimination of parathyroids from the lymph nodes in cases of thyroid malignancy.