Rapid screening of critically ill patients for low plasma vitamin C concentrations using a point-of-care
Sander Rozemeijer1,2,3,4,5, Bob Smit6, Paul W G Elbers7,8,9,10,11
1Department of Intensive Care Medicine, Amsterdam UMC, Location VUmc, De Boelelaan 1117, 1081 HV, Amsterdam, The Netherlands. s.rozemeijer@amsterdamumc.nl.
Background:
Hypovitaminosis C and vitamin C deficiency are common in critically ill patients and associated with organ dysfunction. Low vitamin C status often goes unnoticed because determination is challenging. The static oxidation reduction potential (sORP) reflects the amount of oxidative stress in the blood and is a potential suitable surrogate marker for vitamin C. sORP can be measured rapidly using the RedoxSYS system, a point-of-care device. This study aims to validate a model that estimates plasma vitamin C concentration and to determine the diagnostic accuracy of sORP to discriminate between decreased and higher plasma vitamin C concentrations.
Methods:
Plasma vitamin C concentrations and sORP were measured in a mixed intensive care (IC) population. Our model estimating vitamin C from sORP was validated by assessing its accuracy in two datasets. Receiver operating characteristic (ROC) curves with areas under the curve (AUC) were constructed to show the diagnostic accuracy of sORP to identify and rule out hypovitaminosis C and vitamin C deficiency. Different cut-off values are provided.
Results:
Plasma vitamin C concentration and sORP were measured in 117 samples in dataset 1 and 43 samples in dataset 2. Bias and precision (SD) were 1.3 ± 10.0 µmol/L and 3.9 ± 10.1 µmol/L in dataset 1 and 2, respectively. In patients with low plasma vitamin C concentrations, bias and precision were - 2.6 ± 5.1 µmol/L and - 1.1 ± 5.4 µmol in dataset 1 (n = 40) and 2 (n = 20), respectively. Optimal sORP cut-off values to differentiate hypovitaminosis C and vitamin C deficiency from higher plasma concentrations were found at 114.6 mV (AUC 0.91) and 124.7 mV (AUC 0.93), respectively.
Conclusion:
sORP accurately estimates low plasma vitamin C concentrations and can be used to screen for hypovitaminosis C and vitamin C deficiency in critically ill patients. A validated model and multiple sORP cut-off values are presented for subgroup analysis in clinical trials or usage in clinical practice.