Patient preferences for ambulatory blood pressure monitoring devices: Wrist-type or arm-type?
Wei-Wei Zeng1, Sze Wa Chan2, Brian Tomlinson3,4
1Shenzhen Baoan Women's and Children's Hospital, Jinan University, Shenzhen, Guangdong Province, China.
Background:
Ambulatory blood pressure monitoring (ABPM) is important in evaluating average 24-hour blood pressure (BP) levels, circadian rhythm, sleeping BP and BP variability but many patients are reluctant to use standard ABPM devices.
Methods:
We compared two validated ABPM devices, the BPro tonometric wrist monitor and the A&D TM-2430 oscillometric upper arm monitor, for agreement of recordings and acceptability in 37 hypertensive patients (aged 55±9 years).
Results:
Successful BP measurements were less frequent with the wrist-type than the arm-type device during the sleeping (66.3% vs. 92.9%, P <0.01) and awake periods (56.2% vs. 86.5%, P <0.01). Comparable paired readings showed no significant difference in systolic BP but diastolic BP (DBP) values were higher with the wrist compared to the arm monitor (24-hour 89±13 vs. 85±14 mmHg, P <0.01) with similar differences awake and sleeping. Bland-Altman analysis showed some large discrepancies between individual arm and wrist monitor measurements. More patients found the wrist monitor more comfortable to use than the arm monitor.
Conclusions:
Despite the difference in individual BP measurements and the systematic overestimation of DBP values with the BPro device, wrist monitors with good patient acceptability may be useful to facilitate ABPM in some patients to provide additional information about cardiovascular risk and response to antihypertensive therapies.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)