Photoneutron-induced damage reduction for cardiac implantable electronic devices using neutron-shielding sheets in
Ryo Kakino1, Mitsuhiro Nakamura2, Naonori Hu3
1Division of Medical Physics, Department of Information Technology and Medical Engineering, Human Health Sciences, Graduate School of Medicine, Kyoto University, 53 Shogoin-Kawahara-cho, Sakyo-ku, Kyoto 606-8507, Japan; Department of Radiation Oncology and Image-applied Therapy, Graduate School of Medicine, Kyoto University, 54 Shogoin-Kawahara-cho, Sakyo-ku, Kyoto 606-8507, Japan; Japan Society for the Promotion of Science, 5-3-1 Kojimachi, Chiyoda-ku, Tokyo 102-0083, Japan.
Purpose:
To evaluate damage reduction in cardiac implantable electronic devices (CIEDs) caused by photoneutrons in high-energy X-ray radiotherapy using a neutron-shielding sheet (NSS).
Methods:
The NSS consists of a bolus with a thickness of 1 or 2 cm (Bls1 or Bls2) as a moderator and several absorbers (20%, 50%, or 80% B4C silicone sheet [B4C20, B4C50, or B4C80] or a 40% LiF silicone sheet [LiF40]). First, a linear accelerator (LINAC) with a water-equivalent phantom was modeled in the simulation and measured experimentally. Several NSSs were placed on the phantom, a Eu:LiCaAlF6 scintillator was placed between the phantom and the NSS, and X-rays were irradiated. The relative counts (Cr = counts when placing the NSS or Bls2) were compared between the experiment and simulation. Second, CIED damage was evaluated in the simulation. The relative damage (Dr = damage when placing or not placing the NSS) was compared among all the NSSs. In addition, the γ-ray and leaking X-ray dose from B4C was measured using a dosimetric film. After determining the optimal NSS combination, Dr value analysis was performed by changing the length of one side and the thickness.
Results:
The Cr values of the simulation and experiment agreed within a 30% percentage difference, except for Bare or LiF40-only. The Dr value was reduced by 43% when Bls2 + B4C80 was applied. The photon dose was less than 5 cGy/1500 MU. The Dr values were smaller for the smaller lengths of one side of B4C80 and decreased as the M-layer thickness increased.
Conclusions:
The CIED damage induced by photoneutrons generated by a LINAC was effectively reduced by applying the optimal NSS.