Effect of computed tomography value error on dose calculation in adaptive radiotherapy with Elekta X-ray volume
Takuya Taniguchi1,2, Takanori Hara3, Tomohiro Shimozato4
1Department of Radiation Oncology, Asahi University Hospital, Gifu, Japan.
Purpose:
We evaluated the effect of changing the scan mode of the Elekta X-ray volume imaging cone beam computed tomography (CBCT) on the accuracy of dose calculation, which may be affected by computed tomography (CT) value errors in three dimensions.
Methods:
We used the electron density phantom and measured the CT values in three dimensions. CT values were compared with planning computed tomography (pCT) values for various materials. The evaluated scan modes were for head and neck (S-scan), chest (M-scan), and pelvis (L-scan) with various collimators and filter systems. To evaluate the effects of the CT value error of the CBCT on dose error, Monte Carlo calculations of dosimetry were performed using pCT and CBCT images.
Results:
The L-scan had a CT value error of approximately 800 HU at the isocenter compared with the pCT. Furthermore, inhomogeneity in the longitudinal CT value profile was observed in the bone material. The dose error for ±100 HU difference in CT values for the S-scan and M-scan was within ±2%. The center of the L-scan had a CT error of approximately 800 HU and a dose error of approximately 6%. The dose error of the L-scan occurred in the beam path in the case of both single field and two parallel opposed fields, and the maximum error occurred at the center of the phantom in the case of both the 4-field box and single-arc techniques.
Conclusions:
We demonstrated the three-dimensional CT value characteristics of the CBCT by evaluating the CT value error obtained under various imaging conditions. It was found that the L-scan is considerably affected by not having a unique bowtie filter, and the S-scan without the bowtie filter causes CT value errors in the longitudinal direction. Moreover, the CBCT dose errors for the 4-field box and single-arc irradiation techniques converge to the isocenter.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...