Related Experiment Video
Updated: Oct 24, 2025

Long-Term Continuous Measurement of Renal Blood Flow in Conscious Rats
Published on: February 8, 2022
Reliability of monitoring acid-base and electrolyte parameters through circuit lines during regional citrate
Fang Wang1, Mingjin Dai1, Yuliang Zhao2
1Department of Nephrology, West China Hospital, Sichuan University/West China School of Nursing, Sichuan University, Chengdu, Sichuan, China.
Background:
The current practice involves blood sampling from the circuit line to measure acid-base and electrolyte parameters during continuous renal replacement therapy (CRRT). However, there is limited evidence supporting its reliability due to the effects of anticoagulant mechanism and access recirculation associated with regional citrate anticoagulation (RCA).
Aim:
To evaluate the reliability of monitoring acid-base and electrolyte parameters through circuit lines in regular and reversed connections during RCA-CRRT.
Study Design:
In this prospective cohort study, we included critically ill patients receiving RCA-CRRT via a double-lumen catheter. During the second hour after CRRT initiation, we collected blood samples to monitor acid-base and electrolyte parameters and their levels were compared between samples from the circuit lines (at 0, 3, and 5 minutes) and those from the central venous catheter (CVC) line (at 0 minute). During this time, CRRT switched to the replacement state as controls.
Results:
We observed 128 CRRT circuits in 60 adult patients receiving RCA-CRRT. Ninety-eight (76.6%) circuits had regular connections, while 30 (23.4%) had reversed connections. Among regular connections, no differences were observed in any acid-base or electrolyte parameters between samples from the CVC line and those from the circuit line at all time points (P > .05). Among reversed connections, ionized calcium levels were dramatically decreased at all three time points in samples from the circuit line compared with those from the CVC line (0.65 ± 0.12, 0.72 ± 0.11, and 0.78 ± 0.99 vs 0.98 ± 0.07 mmol/L, P < .001), with comparable levels of other acid-base or electrolyte parameters between the sampling patterns (P > .05).
Conclusions:
Acid-base and electrolyte parameters could be reliably monitored through the circuit line during RCA-CRRT in regular connections. However, in reversed connections, pre-filter ionized calcium concentrations determined through the circuit line were lower than those determined through the CVC line.
Relevance To Clinical Practice:
We suggest sampling from arterial or CVC lines rather than from the circuit line in a reversed connection during RCA-CRRT.
Related Concept Videos
Continuous Renal Replacement Therapy
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Renal Regulation of Acid-Base Balance
In the kidneys, cells within the proximal convoluted tubules (PCT) and the collecting ducts secrete hydrogen ions (H+) into the tubular fluid. Specifically, in the PCT, Na+/H+ antiporters secrete H+ while reabsorbing Na+.
However, the intercalated cells in...
Acute Kidney Injury IV: Diagnostic Studies and Prevention
Acute Kidney Injury V: Interprofessional Care
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...