Analysis of the determining factors of detectable P-wave and amplitude of QRS complex sensed by implantable loop
Jang Ho Ahn1, Hyunho Ryu1, Il-Young Oh1,2
1College of Medicine Seoul National University Seoul South Korea.
Background:
Determining factors for sufficient QRS amplitude and discernible P-wave sensing in implantable loop recorder (ILR) are unknown. We aimed to investigate determining factors and ILR implantation angle that may improve QRS complex and P-wave sensing in ILR.
Methods:
We retrospectively reviewed 220 patients who underwent ILR implantation or follow-up analysis. Patient demographic, clinical, echocardiography, electrocardiography, heart angle, and ILR angle data were collected as predictor variables. Associations between ILR QRS amplitude/P-wave detectability and each predictor variable were investigated.
Results:
Univariate linear regression showed that ILR QRS amplitude was significantly associated with age, height, ILR angle, and QRS amplitudes of 12-lead electrocardiogram (ECG) (lead I, II, aVR [inverted aVR], aVF, V1-V6) and Holter ECG (lead V3, V5). Among discrete variables, only left ventricular hypertrophy (LVH) affected ILR QRS amplitude (P = .016). A multivariate linear regression analysis revealed that ILR angle (β = -0.008, P < .001), lead aVR amplitude (β = 0.469, P = .003), Holter lead V5 amplitude (β = 0.116, P = .049), Age (β = -0.005, P = .014), and LVH (β = 0.213, P = .031) were independent determinants of ILR QRS amplitude. Logistic regression revealed that heart angle significantly affected ILR P-wave detectability (β = 0.12, P = .008). Multiple logistic regression revealed that heart angle (β = 0.121, P = .013) and lead V1 amplitude (β = 28.1, P = .034) were independent determinants of ILR P-wave detectability.
Conclusion:
ILR insertion angle, lead aVR QRS amplitude, Holter lead V5 QRS amplitude, age, and LVH are determinants of ILR QRS amplitude. Heart angle and lead V1 P-wave amplitude of 12-lead ECG are determinants of ILR P-wave detectability.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Electrophysiology of Normal Cardiac Rhythm