Related Experiment Video
Updated: Oct 24, 2025

Simultaneous Laryngopharyngeal and Conventional Esophageal pH Monitoring
Published on: December 14, 2020
Prediction of Esophageal Retention: A Study Comparing High-Resolution Manometry and Functional Luminal Imaging Probe
Dustin A Carlson1, Alexandra J Baumann1, Jacqueline E Prescott1
1Division of Gastroenterology and Hepatology, Department of Medicine, Feinberg School of Medicine, Northwestern University, Chicago, Illinois, USA.
Introduction:
High-resolution manometry (HRM) is generally considered the primary method to evaluate esophageal motility; functional luminal imaging probe (FLIP) panometry represents a novel method to do so and is completed during sedated endoscopy. This study aimed to compare HRM and FLIP panometry in predicting esophageal retention on timed barium esophagram (TBE).
Methods:
A total of 329 adult patients who completed FLIP, HRM, and TBE for primary esophageal motility evaluation were included. An abnormal TBE was defined by a 1-minute column height >5 cm or impaction of a 12.5-mm barium tablet. The integrated relaxation pressure (IRP) on HRM was assessed in the supine and upright patient positions. Esophagogastric junction (EGJ) opening was evaluated with 16-cm FLIP performed during sedated endoscopy through EGJ-distensibility index and maximum EGJ diameter.
Results:
Receiver operating characteristic curves to identify an abnormal TBE demonstrated AUC (95% confidence interval) of 0.79 (0.75-0.84) for supine IRP, 0.79 (0.76-0.86) for upright IRP, 0.84 (0.79-0.88) for EGJ-distensibility index, and 0.88 (0.85-0.92) for maximum EGJ diameter. Logistic regression to predict abnormal TBE showed odds ratios (95% confidence interval) of 1.8 (0.84-3.7) for consistent IRP elevation and 39.7 (16.4-96.2) for reduced EGJ opening on FLIP panometry. Of 40 patients with HRM-FLIP panometry discordance, HRM-IRP was consistent with TBE in 23% while FLIP panometry was consistent with TBE in 78%.
Discussion:
FLIP panometry provided superior detection of esophageal retention over IRP on HRM. However, application of a complementary evaluation involving FLIP panometry, HRM, and TBE may be necessary to accurately diagnose esophageal motility disorders.
Related Concept Videos
Gastroesophageal Reflux Disease II: Clinical Features and Management
Clinical Manifestations
GERD presents itself in a multitude of ways, with symptoms varying from person to person. The hallmark symptoms are...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...