Dosimetry of a novel converging X-ray source for kilovoltage radiotherapy
Tyler Stalbaum1, Larry Partain1, Michael D Weil1
1Imatrex Inc. and Precision RT Inc., Las Vegas, Nevada, USA.
Purpose:
The objective of this work was to evaluate phantom dosimetry of a novel kilovoltage (kV) X-ray source, which employs a stationary tungsten anode and a linearly swept scanning electron beam. The source utilizes converging X-ray collimation along with orthogonal mechanical rotation to distribute surface flux over large area. In this study, this was investigated as a potential solution to fast-falloff limitations expected with kV radiotherapy. This was done with the aim of future clinical development of a lower cost radiotherapy alternative to megavoltage (MV) linac systems.
Methods:
Radiochromic film was employed for dosimetry on the kV X-ray source of the linear-converging radiotherapy system (LCRS). The source utilizes charge particle optics to magnetically deflect and focus an electron beam along a stationary, reflection tungsten target in an ultra-high-vacuum stainless-steel chamber. Resulting X-rays were collimated into converging beamlets that span a large planar angle and converge at the system isocenter. In this study, radiochromic film dosimetry was done at 140 and 145 kVp for a designated planning treatment volume (PTV) of 4 cm diameter. An acrylic phantom was employed for dose distribution measurements of stationary and rotational delivery. Film dosimetry was evaluated in planes parallel to the source X-ray window at various depths, as well as in the plane of gantry rotation.
Results:
At 140 and 145 kVp and using a collimated 4 cm square field at depth, lesion-to-skin dose ratio was shown to improve with additional beams from different relative source positions, where the different beams are focused at the same isocenter and do not overlap at the phantom surface. It was only possible to achieve a 1:1 Dmax -to-surface ratio with four delivery beams, but the ratio improved to 4:1 with 12 beams, focused at the same isocenter depth of 7.8 cm in an acrylic phantom. For the tests conducted, the following Dmax -to-surface ratios were obtained: 0.4:1 lesion-to-skin ratio for stationary delivery from one entry beam, 0.71:1 lesion-to-skin ratio was obtained for two beams, 1.07:1 ratio for four beams, and 4:1 for 12 beams. Dose-depth profiles were evaluated for stationary and rotational dosimetry. Additionally, rotational dosimetry was measured for a case more analogous to a clinical scenario, where the isocenter was located at an off-center simulated lesion.
Conclusions:
The results demonstrate potential dose-depth improvements with kV arc therapy by distributing the surface flux with a wide converging beam along with perpendicular mechanical source rotation of the LCRS. The system delivered tolerable dose to a large surface area when a threshold of multiple, separated beams was reached. The radiochromic film data support the feasibility of the construct of the LCRS kV radiotherapy system design.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Radiation: Applications
The average...