Advanced Closed-Loop Control System Improves Postprandial Glycemic Control Compared With a Hybrid Closed-Loop System
Jose Garcia-Tirado1, Jenny L Diaz1, Rebeca Esquivel-Zuniga2
1Center for Diabetes Technology, University of Virginia, Charlottesville, VA.
Objective:
Meals are a major hurdle to glycemic control in type 1 diabetes (T1D). Our objective was to test a fully automated closed-loop control (CLC) system in the absence of announcement of carbohydrate ingestion among adolescents with T1D, who are known to commonly omit meal announcement.
Research Design And Methods:
Eighteen adolescents with T1D (age 15.6 ± 1.7 years; HbA1c 7.4 ± 1.5%; 9 females/9 males) participated in a randomized crossover clinical trial comparing our legacy hybrid CLC system (Unified Safety System Virginia [USS]-Virginia) with a novel fully automated CLC system (RocketAP) during two 46-h supervised admissions (each with one announced and one unannounced dinner), following 2 weeks of data collection. Primary outcome was the percentage time-in-range 70-180 mg/dL (TIR) following the unannounced meal, with secondary outcomes related to additional continuous glucose monitoring-based metrics.
Results:
Both TIR and time-in-tight-range 70-140 mg/dL (TTR) were significantly higher using RocketAP than using USS-Virginia during the 6 h following the unannounced meal (83% [interquartile range 64-93] vs. 53% [40-71]; P = 0.004 and 49% [41-59] vs. 27% [22-36]; P = 0.002, respectively), primarily driven by reduced time-above-range (TAR >180 mg/dL: 17% [1.3-34] vs. 47% [28-60]), with no increase in time-below-range (TBR <70 mg/dL: 0% median for both). RocketAP also improved control following the announced meal (mean difference TBR: -0.7%, TIR: +7%, TTR: +6%), overall (TIR: +5%, TAR: -5%, TTR: +8%), and overnight (TIR: +7%, TTR: +19%, TAR: -5%). RocketAP delivered less insulin overall (78 ± 23 units vs. 85 ± 20 units, P = 0.01).
Conclusions:
A new fully automated CLC system with automatic prandial dosing was proven to be safe and feasible and outperformed our legacy USS-Virginia in an adolescent population with and without meal announcement.
Related Concept Videos
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
Oral Hypoglycemic Agents: Glinides
Hypoglycemia and Glucagon
Glucagon-like Receptor Agonists
GLP-1, when administered in high doses intravenously, triggers insulin secretion, inhibits glucagon release, slows gastric emptying, reduces food intake, and restores normal insulin secretion. However, its rapid inactivation by...
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...