Microbial growth and importance of flushing inside closed-type infusion devices during administration of lipid

Sachiko Omotani1, Yasutoshi Hatsuda1, Yasuhiro Katsui1

  • 1Faculty of Pharmacy, Osaka Ohtani University.

Insights

Bacterial contamination in lipid emulsions used for Total Parenteral Nutrition (TPN) can lead to microbial growth within infusion devices. Flushing infusion devices with saline post-infusion effectively prevents bacterial proliferation, ensuring patient safety.

Area of Science:

  • Microbiology
  • Clinical Pharmacy
  • Infection Control

Background:

  • Investigated microbial growth in lipid emulsions administered with Total Parenteral Nutrition (TPN).
  • Assessed bacterial proliferation in stagnant lipid emulsions within closed infusion devices.
  • Explored the efficacy of saline flushing to prevent microbial contamination in infusion devices.

Purpose of the Study:

  • To determine the extent of microbial growth in lipid emulsions during TPN.
  • To evaluate the effectiveness of saline flushing in preventing bacterial growth in infusion devices.
  • To identify factors influencing bacterial proliferation and the efficacy of flushing.

Main Methods:

  • Introduced Escherichia coli into lipid emulsion and infused simultaneously with TPN solution.
  • Flushed infusion devices with varying volumes of saline (0, 5, 10, 20 mL) post-infusion.
  • Quantified viable bacteria in stagnant infusion solution collected before and 24 hours after TPN infusion.

Main Results:

  • Significant increase in viable E. coli observed in infusion devices without flushing after 24 hours.
  • Saline flushing effectively prevented bacterial growth in the infusion devices.
  • Bacterial counts varied based on infusion device type, internal structure, and saline flush volume.

Conclusions:

  • E. coli readily multiplies within closed infusion devices when present in lipid emulsions.
  • Post-infusion saline flushing is a critical intervention to prevent microbial growth.
  • Careful management of TPN infusions and infusion devices is essential for infection prevention.