A fully automated high-throughput liquid chromatography tandem mass spectrometry method for measuring creatinine in
Neelam Zahoor1, Uliana Danilenko1, Hubert W Vesper1
1Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, 4770 Buford Hwy NE, MS F25, Atlanta, GA 30341, USA.
Abstract:
Reliable creatinine measurements are important to evaluate kidney function and for creatinine correction to reduce biological variability of other urinary analytes. A high-throughput, accurate liquid chromatography tandem mass spectrometry method for quantitation of human urinary creatinine has been developed and validated. Sample preparation was fully automated including cryovial decapping, sample ID scanning and two serial dilution steps. Quantitation was performed using a stable isotope-labeled internal standard. Multiplexed chromatographic separation of creatinine was achieved within a one-minute analysis and followed by tandem mass spectrometry in positive electrospray ionization mode. The precursor and product ions of creatinine and D3-creatinine were monitored in selected reaction monitoring mode. Method validation results showed reproducibility with within-run precision of 3.59, 3.49 and 2.84% and between-run precision of 4.01, 3.28 and 3.57% for low, medium and high quality control materials prepared from pooled donor urine, respectively. The method showed excellent accuracy with a bias of -1.94%, -0.78% and -1.07% for three levels of certified reference material. The calibration curve was linear throughout a 7.50-300 mg/dL (0.663-26.5 mmol/L) measurement range (R2 = 0.999), with the mean slope of 0.0115 (95%CI, 0.0108-0.0122) and intercept of 0.0027 (95%CI, 0.0003-0.0051). The limit of detection (LOD) of the method was 3.17 mg/dL (0.280 mmol/L). Analytical specificity was achieved by chromatographically separating creatinine from potentially interfering creatine within a one-minute run and monitoring the Quantitation Ion/Confirmation Ion (QI/CI) ratios in samples. A simple, accurate, high-throughput method was successfully developed for measuring creatinine in human urine samples.
More Related Videos
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration