On-line solid-phase extraction of pharmaceutical compounds from wastewater treatment plant samples using restricted
César A Marasco Júnior1, Douglas M Sartore2, Rafaela S Lamarca1
1National Institute for Alternative Technologies of Detection, Toxicological Evaluation and Removal of Micropollutants and Radioactives (INCT-DATREM), Institute of Chemistry, São Paulo State University (UNESP), P.O. Box 355, 14800-060 Araraquara, São Paulo, Brazil.
Abstract:
An on-line solid phase extraction using a lab-made restricted access media (RAM) was developed as sample preparation procedure for determination of the pharmaceutical compounds caffeine (CAF), carbamazepine (CBZ), norfloxacin (NOR), ciprofloxacin (CIP), fluoxetine (FLX) and venlafaxine in wastewater treatment plant samples by liquid chromatography-tandem mass spectrometry (LC-MS/MS). This method is suitable for use in routine of analysis, avoiding cross-contamination and requiring only a small sample volume (50 µL), with minimal handling. The method was validated according to international guidelines. The chromatographic efficiency was evaluated using peak resolution and asymmetry parameters. Carryover was also evaluated, in order to ensure reliability of the analysis and the ability to reuse the cartridge. Satisfactory linearity (r2 > 0.99) was obtained for all the compounds. The intra- and inter-day precision values were lower than 5.79 and 14.1%, respectively. The limits of detection ranged from 0.01 to 3 µg L-1 and the limits of quantification were from 0.1 to 5 µg L-1. The method was applied to 20 environmental wastewater samples, with caffeine being the most widely detected compound, at the highest concentration of 392 µg L-1, while other compounds were detected in fewer samples at lower concentrations (up to 9.60 µg L-1). The lab-made modification is a cheaper option for on-line sample preparation, compared to commercially available on-line SPE cartridges and RAM columns. Moreover, a high-throughput procedure was achieved, with an analysis time of 16 min including sample preparation and chromatographic separation. The same RAM column was applied over 200 injections including method optimization, validation and application in wastewater samples without loss of analytical response.
More Related Videos
05:46Identification of Pharmaceuticals in The Aquatic Environment Using HPLC-ESI-Q-TOF-MS and Elimination of Erythromycin Through Photo-Induced Degradation
Published on: August 1, 2018
09:04Identifying Per- and Polyfluorinated Chemical Species with a Combined Targeted and Non-Targeted-Screening High-Resolution Mass Spectrometry Workflow
Published on: April 18, 2019
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: