Variability in the perception and application of force used in IVC filter retrievals among interventional
T Y Chan1, U Shaikh1, A England2
1Interventional Radiology Department, Royal Liverpool University Hospital, Prescot St, L78XP, Liverpool, United Kingdom.
Introduction:
A long dwell time is associated with higher inferior vena cava (IVC) filter retrieval complication rates. Manufacturers advise that "excessive force should not be used" during filter retrieval; however, the term "excessive" is subjective and is likely to lead to variability amongst operators. The aims of this study were to 1) ascertain what interventional radiologists consider to be excessive force during filter retrieval and 2) to understand the variability in interventional radiologists' perception of force.
Methods:
The authors recruited interventional radiologists to perform a benchtop simulated filter retrieval. Participants were invited to pull on a modified force tester attached to a Gunther Tulip filter retrieval set (GTRS). The participants were asked to pull as if they were retrieving an IVC filter and stop when they felt it was clinically unsafe to apply greater force. They were then asked to replicate forces of 10N and 50N, respectively. Each of the three tasks was completed three times. Data were obtained on the clinical experience of the participants with specific questions focusing on their filter retrieval practices.
Results:
The range of maximum forces applied during filter retrieval varied between 0.8 and 79.8N. When asked to replicate 10N and 50N, for attempt-1, the median forces produced were 23.5N and 38.1N, respectively. A trend analysis showed that those who overestimate 10N are more likely to apply a greater overall maximum force (rs = 0.622; P < 0.001).
Conclusion:
There is wide variation in what interventional radiologists consider to be the maximum safe force to apply during IVC filter retrieval.
Implications For Practice:
Manufacturers and operators should consider methods in which only a safe range of forces can be applied during an IVC filter retrieval. Operators may wish to undertake 'personal' force calibration as part of training in interventional radiology.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
06:50Troubleshooting FoCUS Image Acquisition: Patient Positioning, Transducer Manipulation, and Image Optimization
Published on: March 3, 2023
Related Concept Videos
Imaging Studies VII: Vascular Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Varicose Veins II: Diagnostic Studies and Interprofessional Care
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies II: Ultrasonography