Innovation in measuring obese patients' blood pressure: measurement with conical wrapping technique
Seda Şahan1, Sevil Şahin2, Hakan Aygün3
1Nursing Department, Izmir Bakircay University, Health Sciences Faculty, İzmir.
Introduction:
Accurate measurement of blood pressure (BP) is difficult in obese patients due to their upper arm shape and width. Errors made during BP measurement cause misdiagnosis or lead to wrong treatment.
Method:
The data of this methodological study were collected from 40 patients who were 18 years old or above, with a BMI above 25 kg/m2, and who were followed up with radial catheter arterial pressure in adult intensive care and postoperative care units between March and December 2020.
Results:
In total 55% of the participants were male with a mean age of 63.9 years, mean height of 165.7 cm, mean weight of 85.2 kg and mean BMI of 31.0 ± 2.1 kg/m2. Intra-arterial BP (IABP) measurement values were found to be correlated with conical wrapping technique and cylindrical wrapping technique (P < 0.001). As a result, a statistically significance was found between conical SBP and intra-arterial SBP (r = 0.921; P < 0.05) and conical DBP and intra-arterial DBP (r = 0.902; P < 0.05).
Conclusion:
It was determined that the SBP and DBP results of the conical wrapping technique were closer to the results obtained by IABP measurements. For this reason, it is more appropriate to measure BP with conical wrapping technique in clinically obese patients.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
08:54Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
Published on: May 15, 2009
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: