High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Clinical implications of compromised 82Rb PET data acquisition
Andrew Van Tosh1,2, J Jane Cao1, John R Votaw3
1St. Francis Hospital, Roslyn, NY, USA.
Background:
We wished to document the prevalence and quantitative effects of compromised 82Rb PET data acquisitions on myocardial flow reserve (MFR).
Methods And Results:
Data were analyzed retrospectively for 246 rest and regadenoson-stress studies of 123 patients evaluated for known or suspected CAD. An automated injector delivered pre-determined activities of 82Rb. Automated quality assurance algorithms identified technical problems for 7% (9/123) of patients. Stress data exhibited 2 instances of scanner saturation, 1 blood peak detection, 1 blood peak width, 1 gradual patient motion, and 2 abrupt patient motion problems. Rest data showed 1 instance of blood peak width and 2 abrupt patient motion problems. MFR was lower for patients with technical problems flagged by the quality assurance algorithms than those without technical problems (1.5 ± 0.5 versus 2.1 ± 0.7, P = 0.01), even though rest and stress ejection fraction, asynchrony and relative myocardial perfusion measures were similar for these two groups (P > 0.05), suggesting that MFR accuracy was adversely affected by technical errors.
Conclusion:
It is important to verify integrity of 82Rb data to ensure MFR computation quality.
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019