Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Coronary Compression Following Percutaneous Pulmonary Valve Implantation Despite a Negative Balloon Sizing Test:
Maite Velázquez Martín1, Fernando Sarnago Cebada2, Nicolás Maneiro Melón2
1Cardiology Department, University Hospital 12 de Octubre, Madrid, Spain; Instituto de Investigación Sanitaria Hospital 12 de Octubre (imas12), Spain; CIBERCV, Spain.
Abstract:
A 16-years-old male with prior diagnosis of situs inversus totalis and pulmonary atresia with interventricular communication underwent percutaneous pulmonary valve implantation 3 months after successful RVOT stenting following a negative balloon sizing test. Once finished the procedure, after consciousness recovery in the intensive care unit, the patient developed oppressive chest pain with very subtle electrocardiographic changes over his basal right bundle branch block. An urgent coronary angiography showed a severe stenosis in the proximal right coronary artery with TIMI 2 distal flow. Intravascular ultrasound imaging confirmed extrinsic compression. A 4 × 21 mm drug-eluting stent was successfully implanted relieving symptoms immediately. TIMI 3 flow was restored and good apposition and expansion were confirmed with intravascular ultrasound. A carefully review of the procedure showed that the pitfall responsible for this complication was the oversizing of the valve with respect to the size of the balloon used for the sizing test.
More Related Videos
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
06:59Catheter-based Endovascular Angioplasty for Fibrosing Mediastinitis-associated Pulmonary Vein Stenosis
Published on: August 26, 2025