Helmut Neumann1,2, Andreas Kreft3, Visvakanth Sivanathan1
1Department of Interdisciplinary Endoscopy, I. Medizinische Klinik und Poliklinik, University Hospital, Mainz, Germany.
Background:
Linked color imaging (LCI) has been shown to be effective in multiple randomized controlled trials for enhanced colorectal polyp detection. Recently, artificial intelligence (AI) with deep learning through convolutional neural networks has dramatically improved and is increasingly recognized as a promising new technique for enhancing colorectal polyp detection.
Aim:
This study aims to evaluate a newly developed computer-aided detection (CAD) system in combination with LCI for colorectal polyp detection.
Methods:
First, a convolutional neural network was trained for colorectal polyp detection in combination with the LCI technique using a dataset of anonymized endoscopy videos. For validation, 240 polyps within fully recorded endoscopy videos in LCI mode, covering the entire spectrum of adenomatous histology, were used. Sensitivity (true-positive rate per lesion) and false-positive frames in a full procedure were assessed.
Results:
The new CAD system used in LCI mode could process at least 60 frames per second, allowing for real-time video analysis. Sensitivity (true-positive rate per lesion) was 100%, with no lesion being missed. The calculated false-positive frame rate was 0.001%. Among the 240 polyps, 34 were sessile serrated lesions. The detection rate for sessile serrated lesions with the CAD system used in LCI mode was 100%.
Conclusions:
The new CAD system used in LCI mode achieved a 100% sensitivity per lesion and a negligible false-positive frame rate. Note that the new CAD system used in LCI mode also specifically allowed for detection of serrated lesions in all cases. Accordingly, the AI algorithm introduced here for the first time has the potential to dramatically improve the quality of colonoscopy.
More Related Videos
07:12Spatio-Temporal In Vivo Imaging of Ocular Drug Delivery Systems using Fiberoptic Confocal Laser Microendoscopy
Published on: September 27, 2021
09:24Optical Screening of Novel Bacteria-specific Probes on Ex Vivo Human Lung Tissue by Confocal Laser Endomicroscopy
Published on: November 29, 2017
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures II: Colonoscopy