Achieving high-resolution 1H-MRSI of the human brain with compressed-sensing and low-rank reconstruction at 7 Tesla
Antoine Klauser1, Bernhard Strasser2, Bijaya Thapa2
1Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Boston, MA, United States; Department of Radiology and Medical Informatics, University of Geneva, Switzerland; Center for Biomedical Imaging (CIBM), Geneva, Switzerland.
Abstract:
Low sensitivity MR techniques such as magnetic resonance spectroscopic imaging (MRSI) greatly benefit from the gain in signal-to-noise provided by ultra-high field MR. High-resolution and whole-slab brain MRSI remains however very challenging due to lengthy acquisition, low signal, lipid contamination and field inhomogeneity. In this study, we propose an acquisition-reconstruction scheme that combines 1H free-induction-decay (FID)-MRSI sequence, short TR acquisition, compressed sensing acceleration and low-rank modeling with total-generalized-variation constraint to achieve metabolite imaging in two and three dimensions at 7 Tesla. The resulting images and volumes reveal highly detailed distributions that are specific to each metabolite and follow the underlying brain anatomy. The MRSI method was validated in a high-resolution phantom containing fine metabolite structures, and in five healthy volunteers. This new application of compressed sensing acceleration paves the way for high-resolution MRSI in clinical setting with acquisition times of 5 min for 2D MRSI at 2.5 mm and of 20 min for 3D MRSI at 3.3 mm isotropic.