In Vitro Thrombosis Test for Ventricular Assist Devices
Published on: March 21, 2025
Safety Evaluation of a Forward-Viewing Intravascular Transducer for Sonothrombolysis: An in Vitro and ex Vivo Study
Leela Goel1, Huaiyu Wu2, Bohua Zhang2
1Department of Mechanical and Aerospace Engineering, North Carolina State University, Raleigh, North Carolina, USA; Joint Department of Biomedical Engineering, University of North Carolina at Chapel Hill and North Carolina State University, Raleigh, North Carolina, USA.
Abstract:
Recent in vitro work has revealed that a forward-viewing intravascular (FVI) transducer has sonothrombolysis applications. However, the safety of this device has yet to be evaluated. In this study, we evaluated the safety of this device in terms of tissue heating, vessel damage and particle debris size during sonothrombolysis using microbubbles or nanodroplets with tissue plasminogen activator, in both retracted and unretracted blood clots. The in vitro and ex vivo sonothrombolysis tests using FVI transducers revealed a temperature rise of less than 1°C, no vessel damage as assessed by histology and no downstream clot particles >500 µm. These in vitro and ex vivo results indicate that the FVI transducer poses minimal risk for sonothrombolysis applications and should be further evaluated in animal models.