Related Experiment Video
Updated: Oct 22, 2025

Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
Analysis of Surface Roughness and Three-dimensional Scanning Topography of Zirconia Implants before and after
R Arun Jaikumar1, Suma Karthigeyan2, T R Ramesh Bhat1
1Department of Prosthodontics, Best Dental Science College, Madurai, Tamil Nadu, India.
Aim:
To analyze surface roughness and three-dimensional (3D) scanning topography parameters of zirconia implants before and after photofunctionalization by atomic force microscopy (AFM).
Materials And Methods:
Ten commercially available zirconia implants five each in the study and control group were taken. The study group was subjected to ultraviolet (UV) radiation for 48 h using the shorter wavelength of 254 nm. After washing all the implants with 70% alcohol and drying, 3D surface topography and roughness parameters were analyzed using CSC 17 probe AFM at three different magnifications 25 μm, 50 μm, and 80 μm, respectively.
Results:
The surface topography and calculated mean amplitude, spatial, and hybrid parameters of the study group were higher than the control group (P < 0.05) in all three magnifications. Up to scale depth and peak value for the study and control group were (-0.4-0.4: -2-1) (-0.75 to 0.6:-1-1.3) (-0.75--0.5: -1.5-1.3) for the study and control group at 25, 50, and 80 μm magnification, respectively. This indicates that photofunctionalization increased surface roughness of Zirconia implants to desirable extent.
Conclusion:
There is a definite difference in the quantitative topographic characteristics with zirconia implants being microroughned after photofunctionalization (UV treatment).