Residual volume fraction during walking using wireless air plethysmography in patients with chronic deep venous
Oscar Maleti1, Andrew Nicolaides2, Sara Guerzoni1
1Division of Vascular Surgery, Department of Cardiovascular Surgery, Hesperia Hospital, Modena, Italy.
Objective:
Traditional air plethysmography (APG) provides a quantitative measure of the residual volume fraction (RVF) after 10 tiptoe movements. The recent development of a wireless Bluetooth (Bluetooth SIG, Inc, Kirkland, Wash) APG device, the PicoFlow (Microlab Elettronica, Padua, Italy), enabled us to measure RVF during normal walking. The aim of our study was to compare the RVF obtained during tiptoeing with RVF obtained during normal walking in patients with deep venous pathology (ie, reflux and/or obstruction).
Methods:
A total of 61 consecutive symptomatic patients (27 women and 34 men; median age, 46 years; range, 18-79 years) with chronic venous disease due to deep venous pathology (venous reflux or obstruction, or both) before treatment or persisting after intervention were included in the present study. Of the 122 total limbs examined, 79 were affected by deep chronic venous disease and 43 contralateral limbs were normal with normal deep veins and acted as controls. The APG examination was performed using the PicoFlow device using the standard examination technique. The RVF was calculated from the residual volume at the end of 10 tiptoe movements and also during normal walking.
Results:
At the end of the 10 tiptoe movements, the mean ± standard deviation RVF was 27.0% ± 13.2% in the limbs with normal deep veins and 38.8% ± 16.9% in the limbs with deep chronic venous disease (P < .001). During walking, when a steady state in volume was reached, the RVF was 26.3% ± 17.8% in the limbs with normal deep veins and 43.1% ± 18.6% in limbs with deep venous disease (P < 0.001). A significant difference was found between limbs with normal deep veins and limbs with deep venous reflux, irrespective of which exercise was performed. However, the mean RVF between the limbs with normal deep veins and those with outflow obstruction in the absence of reflux was significant during walking (P = .012) but not during tiptoeing (P = .212). The mean RVF was higher in the C3 to C6 limbs than in the C0 to C2 limbs with tiptoeing (29.9% ± 14.5% vs 38.3% ± 17.0%; P < .006). Similar results were obtained with walking (29.2% ± 18.0% vs 42.4% ± 18.8%; P < .004).
Conclusions:
In limbs with normal deep veins and deep veins with reflux, the RVF measured during walking with wireless APG was similar to the RVF obtained during tiptoeing. However, in the limbs with outflow obstruction in the absence of reflux, the RVF during walking was higher than the RVF after tiptoeing. Our results have shown that the evaluation of RVF during walking is feasible and practical.
More Related Videos
08:05Occlusion of the Great and Small Saphenous Vein Using Copolymeric Glue Based on N-Butyl Cyanoacrylate and Methacryloxy Sulfolane
Published on: December 9, 2022
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system...
Venous Thrombosis III: Interprofessional Care
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: