Bioreactance reliably detects preload responsiveness by the end-expiratory occlusion test when averaging and refresh
Francesco Gavelli1,2, Alexandra Beurton3, Jean-Louis Teboul3
1Service de Médecine Intensive-Réanimation, Université Paris-Saclay, AP-HP, Hôpital de Bicêtre, DMU CORREVE, Inserm UMR S_999, FHU SEPSIS, CARMAS, 78, Rue du Général Leclerc, 94 270, Le Kremlin-Bicêtre, France. francesco.gavelli@uniupo.it.
Background:
The end-expiratory occlusion (EEXPO) test detects preload responsiveness, but it is 15 s long and induces small changes in cardiac index (CI). It is doubtful whether the Starling bioreactance device, which averages CI over 24 s and refreshes the displayed value every 4 s (Starling-24.4), can detect the EEXPO-induced changes in CI (ΔCI). Our primary goal was to test whether this Starling device version detects preload responsiveness through EEXPO. We also tested whether shortening the averaging and refresh times to 8 s and one second, respectively, (Starling-8.1) improves the accuracy of the device in detecting preload responsiveness using EEXPO.
Methods:
In 42 mechanically ventilated patients, during a 15-s EEXPO, we measured ∆CI through calibrated pulse contour analysis (CIpulse, PiCCO2 device) and using the Starling device. For the latter, we considered both CIStarling-24.4 from the commercial version and CIStarling-8.1 derived from the raw data. For relative ∆CIStarling-24.4 and ∆CIStarling-8.1 during EEXPO, we calculated the area under the receiver operating characteristic curve (AUROC) to detect preload responsiveness, defined as an increase in CIpulse ≥ 10% during passive leg raising (PLR). For both methods, the correlation coefficient vs. ∆CIpulse was calculated.
Results:
Twenty-six patients were preload responders and sixteen non preload-responders. The AUROC for ∆CIStarling-24.4 was significantly lower compared to ∆CIStarling-8.1 (0.680 ± 0.086 vs. 0.899 ± 0.049, respectively; p = 0.027). A significant correlation was observed between ∆CIStarling-8.1 and ∆CIpulse (r = 0.42; p = 0.009), but not between ∆CIStarling-24.4 and ∆CIpulse. During PLR, both ∆CIStarling-24.4 and ∆CIStarling-8.1 reliably detected preload responsiveness.
Conclusions:
Shortening the averaging and refresh times of the bioreactance signal to 8 s and one second, respectively, increases the reliability of the Starling device in detection of EEXPO-induced ∆CI.
Trial Registration:
No.
Idrcb:
2018-A02825-50. Registered 13 December 2018.
More Related Videos
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: