Calibrated automated thrombogram II: removing barriers for thrombin generation measurements
P L A Giesen1, A J W Gulpen2, R van Oerle2
1CoagScope BV, Maastricht, The Netherlands.
Thrombosis Journal
|August 29, 2021
Summary
A new thrombin generation assay (CAT-II) improves monitoring of direct thrombin inhibitors. This modified method accurately detects anticoagulant effects, unlike the standard assay (CAT-I), enhancing patient care.
Area of Science:
- Hemostasis and Thrombosis Research
- Clinical Chemistry and Diagnostics
- Pharmacodynamics of Anticoagulants
Background:
- Thrombin generation (TG) via Calibrated Automated Thrombogram (CAT-I) assesses coagulation balance but has limitations.
- CAT-I struggles with substrate consumption and cannot calibrate with direct thrombin inhibitors (DTIs).
- A modified assay, CAT-II, was developed to address these limitations in TG assessment.
Purpose of the Study:
- To compare the performance of the modified TG assay (CAT-II) against the standard CAT-I method.
- To evaluate CAT-II's efficacy with various anticoagulants and interfering substances.
- To assess a new automated TG instrument (ST®-Genesia) using a standardized calibration method.
Main Methods:
- Thrombin generation was measured using CAT-I, CAT-II, and ST®-Genesia in plasma samples.
- Samples included controls, those with dabigatran, rivaroxaban, apixaban, and interfering substances.
- CAT-II and ST®-Genesia utilized a novel calibration method with standard thrombin and AMC in separate plasma.
Main Results:
- CAT-II showed higher TG values than CAT-I in anticoagulant-free plasmas; both methods had <5% intra-assay variation.
- CAT-II accurately detected dose-dependent inhibition by dabigatran, a feat missed by CAT-I.
- Both CAT-II and ST®-Genesia effectively corrected for interfering substances and detected oral anticoagulant effects.
Conclusions:
- The CAT-II method demonstrates high similarity to CAT-I in non-DTI samples.
- CAT-II and ST®-Genesia with the new calibration method reliably detect effects of all oral anticoagulants.
- The improved calibration method in CAT-II significantly enhances monitoring of patients on direct thrombin inhibitors.