Determination of Diagnostic Reference Level (DRL) in Common Computed Tomography Examinations with the Modified
Jalal Tabesh1, Maziyar Mahdavi2,3, Gholamhasan Haddadi2,3
1MSc, Department of Radiology, School of Paramedical Sciences, Shiraz University of Medical Sciences, Shiraz, Iran.
Background:
The diagnostic reference level (DRL) is measured with different methods in the common Computed tomography (CT) exams, but it has not been measured through the size-specific dose estimate (SSDE) method in Iran, yet.
Objective:
This study aimed to calculate the local DRL (LDRL) using the new quality control-based dose survey method (QC) and patients' effective diameter (MQC) and compare them with a data collection method (DC) as well as local national DRLs (NDRL).
Material And Methods:
In this cross-sectional study, LDRL, based on the third quartile of volumetric computed tomography dose index (CTDIvol) and dose length product (DLP) values, was calculated for the four common CT examinations in four CT scan centers affiliated with Shiraz University of Medical Sciences by DC, QC and MQC methods. The CTDIvol of each patient for each CT exam calculated with three methods was compared with paired t-test. Also, the LDRL using MQC method was compared with other national DRL studies.
Results:
There was a significant difference between the CTDIvol values calculated with MQC and QC in all four examinations (P <0.001). The LDRL based on CTDIvol obtained by the MQC method for head, sinus, chest, abdomen, and pelvis were (50, 18, 15, 19) mGy, respectively, and the calculated DLP values were also (735, 232, 519, 984) mGy.cm.
Conclusion:
In MQC, LDRL based on CTDIvol was calculated with a mean difference percentage of (19.2 ± 11.6)% and (27.1 ± 8.1)% as compared to the QC and DC methods, respectively. This difference resulted from the use of the SSDE method and dose accuracy in the QC dose survey.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET