Development and evaluation of a synthetic opioid targeted gas chromatography mass spectrometry (GC-MS) method
Edward Sisco1, Amber Burns2, Arun S Moorthy1
1National Institute of Standards and Technology, Gaithersburg, MD, USA.
Abstract:
As seized drug casework becomes increasingly complex due to the continued prevalence of emerging drugs, laboratories are often looking for new analytical approaches including developing methods for the analysis of specific compounds classes. Recent efforts have focused on the development of targeted gas chromatography mass spectrometry (GC-MS) confirmation methods to compliment the information-rich screening results produced by techniques like direct analysis in real time mass spectrometry (DART-MS). In this work, a method for the confirmation of synthetic opioids and related compounds was developed and evaluated. An 11-component test solution was used to develop a method that focused on minimizing overlapping retention time acceptance windows and understanding the influence of instrument parameters on reproducibility and sensitivity. Investigated settings included column type, flow rate, temperature program, inlet temperature, source temperature, and tune type. Using a DB-200 column, a 35-min temperature ramped method was created. It was evaluated against a suite of 222 synthetic opioids and related compounds, and successfully differentiated all but four compound pairs based on nonoverlapping retention time acceptance windows or objectively different mass spectra. Compared to a general confirmatory method used in casework, the targeted method was up to 25 times more sensitive and provided at least a two-fold increase in retention time differences. Analysis of extracts from actual case samples successfully demonstrated utility of the method and showed no instance of carryover, although the high polarity column required wider retention time windows than other columns.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
05:31Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...