Related Experiment Video
Updated: Oct 22, 2025

Evaluating Regional Pulmonary Deposition using Patient-Specific 3D Printed Lung Models
Published on: November 11, 2020
Using Filters to Estimate Regional Lung Deposition with Dry Powder Inhalers
Scott Tavernini1,2, Dino J Farina3, Andrew R Martin1,2
1Department of Mechanical Engineering, University of Alberta, Edmonton, Alberta, Canada.
Purpose:
To develop an in vitro method to rapidly evaluate regional lung doses delivered by pharmaceutical inhalers. Currently, cascade impactor measurements are used, but these are resource intensive and require significant post processing of in vitro data to arrive at regional deposition estimates.
Methods:
We present a specialized filter apparatus that mimics tracheobronchial (TB) deposition of pharmaceutical aerosols emitted by commercially available dry powder inhalers (DPIs). The filter housing includes an electrostatic neutralizer to eliminate artificial electrostatic filtration effects. Regional deposition (tracheobronchial and alveolar) for four DPIs (Onbrez Breezhaler, Flovent Diskus, Pulmicort Turbuhaler, and Asmanex Twisthaler) was estimated using cascade impactor measurements and an in silico regional deposition model. These estimates were compared to direct measurements of regional deposition as provided by the TB filter mimic and an absolute filter placed downstream of the TB filter housing, representing the alveolar dose.
Results:
The two methods were shown to provide similar estimates of extrathoracic, tracheobronchial, and alveolar deposition, as well as total recovery of active pharmaceutical ingredients.
Conclusions:
Because of its design, the TB filter apparatus makes it possible to estimate regional deposition with inhalers directly using variable inhalation profiles without any additional equipment or changes to the experimental configuration. This method may be useful to expedite development of both innovative and generic drug products as it provides regional respiratory tract deposition estimates using fewer resources than exisiting methods.
More Related Videos
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
04:59Author Spotlight: Developing a Disposable Dosator for Preclinical Testing of Dry Powder Inhalers in Small Animal Models
Published on: August 18, 2023