Related Experiment Video
Updated: Aug 10, 2026

Utilizing Transcranial Magnetic Stimulation to Study the Human Neuromuscular System
Published on: January 20, 2012
A compact 1200 V, 700 A, IGBT-based pulse generator for repetitive transcranial magnetic stimulation in vivo
Daniel Senda1, Haley Strong2, Dustin Hines2
1Department of Electrical and Computer Engineering, University of Nevada, Las Vegas, 4505 S. Maryland Pkwy, Las Vegas, Nevada 89154, USA.
Abstract:
An insulated-gate bipolar transistor (IGBT) pulse generator for repetitive transcranial magnetic stimulation used for in vivo laboratory experiments on small animals, such as mice, is reported. The pulse generator is based upon an IGBT that can switch 700 A of current for 1 ms and that has a DC breakdown voltage of 1200 V. The duration of the design's output pulse is controlled by, and follows, an input trigger pulse. The voltage amplitude of the output pulses is determined by an external high-voltage power supply and the energy stored in a 330 µF capacitor bank. The approach enables the amplitude of the voltage applied across the coil, the length of time the voltage is applied, and the number of times the voltage pulses are applied all to be controlled and adjusted to facilitate a wide range of experimental options. This paper provides a detailed schematic of the design, design discussions, and some representative experimental results. Additionally, the reported design can be scaled to higher currents by using an IGBT with a higher current rating.