Published on: March 24, 2020
Validation of Modified Hertel Exophthalmometer
Rizwana Fathima Jamal1, Emmanuel Azariah1, Deepak Pandyan1
1Department of Oral and Maxillofacial Surgery, Faculty of Dental Sciences, Sri Ramachandra Institute of Higher Education and Research, Chennai, Tamil Nadu, India.
Study Design:
This is a correlation study.
Objective:
The objective was to obtain normative values from modified Hertel exophthalmometer, to compare difference in values between modified and standard Hertel exophthalmometer and to statistically analyze for correlation.
Methods:
The study to validate the modification done to Hertel exophthalmometer was performed at the Department of Oral and Maxillofacial Surgery at our institute. Informed consent was obtained from patients; measurement in 100 volunteers was made with Hertel and modified Hertel exophthalmometer by a single observer, and the measurements taken by the 2 devices were tabulated along with variables such as age and gender. Data were collected and statistical analysis was performed to compare and correlate the values between standard and modified exophthalmometer. Differences of ocular protrusion values in gender and age were also calculated.
Results:
There has been no statistically significant difference in the normative data between standard Hertel and modified Hertel exophthalmometer and in relation to age and gender.
Conclusions:
The modification of the Hertel exophthalmometer is validated and can also be used for relative exophthalmometry to assess the need for treatment and to evaluate the globe position and orbital volume intraoperatively. The proposed modification of the instrument allows it to be used in cases when one of the lateral orbital walls is fractured or displaced due to trauma.
Related Concept Videos
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...