Modeling the Impact of Electrode/Tissue Geometry on Electrical Stimulation in Stereo-EEG
Alexis C Shindhelm1, Brandon J Thio2, Saurabh R Sinha1
1Department of Neurology, Duke University Medical Center, Durham, North Carolina; and.
Purpose:
Electrical stimulation through depth electrodes is used to map function and seizure onset during stereoelectroencephalography in patients undergoing evaluation for epilepsy surgery. Factors such as electrode design, location, and orientation are expected to impact effects of electrical stimulation.
Methods:
We developed a steady-state finite element model of brain tissue including five layers (skull through white matter) and an implanted electrode to explore the impact of electrode design and placement on the activation of brain tissue by electrical stimulation. We calculated electric potentials, current densities, and volume of tissue activated ( Volact ) in response to constant current bipolar stimulation. We modeled two depth electrode designs (3.5- and 4.43-mm intercontact spacing) and varied electrode location and orientation.
Results:
The electrode with greater intercontact spacing produced 8% to 23% larger Volact (1% to 16% considering only gray matter). Vertical displacement of the electrodes by half intercontact space increased Volact for upward displacement (6% to 83% for all brain tissue or -5% to 96% gray matter only) and decreased Volact (1% to 16% or 24% to 49% gray matter only) for downward displacement. Rotating the electrode in the tissue by 30° to 60° with respect to the vertical axis increased Volact by 30% to 90% (20%-48% gray matter only).
Conclusions:
Location and orientation of depth electrodes with respect to surrounding brain tissue have a large impact on the amount of tissue activated during electrical stimulation mapping in stereoelectroencephalography. Electrode design has an impact, although modest for commonly used designs. Individualization of stimulation intensity at each location remains critical, especially for avoiding false-negative results.