Effect of Hypoproteinemia on Electrolyte Measurement by Direct and Indirect Ion Selective Electrode Methods
Mukesh Gohel1, Jemil S Makadia1, Chandan Chakrabarti1
1Department of Biochemistry, Smt NHL Municipal Medical College, Ahmedabad, Gujarat, India.
Abstract:
Objective The aim of this study was to see the effect of hypoproteinemia on electrolyte measurement by two different techniques, that is, direct ion selective electrode (ISE) and indirect ISE. Material and Method It was an observational study in which 90 serum samples with normal protein content (Group-1) were subjected to sodium (Na + ) and potassium (K + ) measurements by direct and indirect ISE methods. In the same way, 90 serum samples with total protein < 5 g/dL (Group-2) were subjected to Na + and K + measurements by direct and indirect ISE methods. Result In samples from Group-1 patients, average Na + was 138.1 ± 4.764 mmol/L by direct ISE method and 139.3 ± 3.887 mmol/L by indirect ISE method while average K + was 4.41 ± 0.644 mmol/L by direct ISE method and 4.40 ± 0.592 mmol/L by indirect ISE method. There was no statistically significant difference in Na + and K + values measured by different methods. In samples from Group-2 patients, measured value of Na + by direct ISE and indirect ISE was 134.57 ± 5.520 mmol/L and 138.64 ± 5.401 mmol/L, respectively. Difference between these two values was statistically significant with p -value of < 0.0001, but direct ISE and indirect ISE measured values of K + was 4.146 ± 0.9639 mmol/L and 4.186 ± 0.8989, respectively, with no significant difference. Conclusion Direct and indirect ISE methods are not comparable and showing significantly different results for Na + in case of hypoproteinemia. So, it is recommended that setups like intensive care unit or emergency department, where electrolyte values have significant treatment outcome, should follow direct ISE method and should compare its previous result with the same method. Both the methods should not be used interchangeably.
More Related Videos
Related Concept Videos
EDTA: Indirect and Alkalimetric Titration
Potentiometry: Membrane Electrodes
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Protein-Drug Binding: Determination Methods
Indirect methods involve isolating the bound drug from its free form in biological samples such as blood, serum, or plasma. These techniques aim to measure the percentage of drugs bound to proteins. Equilibrium dialysis is a commonly used method where the free drug concentration at equilibrium is measured by separating the bound...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...