Dosimetric evaluation for temporomandibular joint cone beam computed tomography exams using different field of view
Carla Alves Pinto1, Gabriela Franco da Rosa Caetano1, Marcos Ely Almeida Andrade2
1Division of Oral Radiology, Faculdade São Leopoldo Mandic, Instituto de Pesquisas São Leopoldo Mandic, Campinas, São Paulo (SP), 13045-755, Brazil.
Abstract:
Objectives.To optimize the absorbed organ dose in relation to the field of view for temporomandibular joint examinations in four cone beam computed tomography devices.Methods.An anthropomorphic adult head and neck phantom, and 192 LiF dosimeters (TLD-100) were used. The dosimeters were placed in the region corresponding to the lens, parotid glands, submandibular glands, and thyroid. Small, medium and large FOVs were selected on Orthopantomograph OP300 Maxio, PaX-i3D Smart, ORTHOPHOS XG, and i-CAT Next Generation device when it was possible.Results.A wide range of absorbed dose values was recorded for all organs due to the different exposure parameters of each device. The radiosensitive organ with the highest dose was the parotid glands. The devices with 5 × 5 cm FOV recorded a lower dose in this protocol, while for the device without a small FOV (≤5 × 5 cm), the lowest dose was observed with the large FOV (6 × 16 cm).Conclusions.We recommend a double exposure with an FOV of 5 × 5 cm in the OP300 Maxio, PaX-i3D Smart, and ORTHOPHOS XG device, while in the i-CAT Next Generation device, a single exposure FOV of 6 × 16 cm is indicated.
More Related Videos
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
09:10Digital Hybrid Model Preparation for Virtual Planning of Reconstructive Dentoalveolar Surgical Procedures
Published on: August 5, 2021
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET