Laboratory evaluation of performance of pulse oximeters from six different manufacturers during motion artifacts
Munirathinam Ganesh Kumar1, Sukhbir Kaur2, Raj Kumar3
1Department of Anaesthesia & Intensive Care, Postgraduate Institute of Medical Education & Research, Chandigarh, 160012, India. mganeshkumar19@gmail.com.
Abstract:
We performed laboratory evaluation of six pulse oximeters from different manufacturers using the Fluke 2XL SpO2 simulator. The pulse oximeter probes were labeled 1 through 6 and tested using the two pre-programed preset functions of the Fluke 2XL SpO2 simulator, level 01 and level 02, for their performance in the presence of motion artifacts. The pulse oximeters were also tested at low perfusion index (PI) settings. The accuracy of the individual probes was ranked as good, mediocre, or poor based on the degree of deviation of the measured SpO2 and pulse rate (PR) by the probes from that generated by the simulator in both the presence and absence of motion artifacts and at lower PI. In preset level 01, probes numbered 1, 2, and 6 performed flawlessly, probes numbered 3, 4, and 5 performed poorly. In preset level 02, all probes performed well in the absence of motion artifacts. When probes were attached in the absence of motion, but readings were recorded after the emergence of motion artifacts, probes 1, 2, and 6 continued to perform well. However, the performance of probes 3, 4, and 5 deteriorated. When probes were attached directly in the presence of motion artifacts, probes 2 and 6 performed well, whereas all other probes performed poorly. Successively lowering the PI degraded performance of probes 3, 4, and 5 at extremely low PI. It is observed that during motion and/or low PI conditions, multiple probes see deterioration in performance.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
07:20A Novel Approach to Overcome Movement Artifact When Using a Laser Speckle Contrast Imaging System for Alternating Speeds of Blood Microcirculation
Published on: August 30, 2017
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...