Related Experiment Video
Updated: Oct 21, 2025

In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Chronic Wear-Induced Bearing Dislocation in a Fixed-Bearing Unicompartmental Knee Arthroplasty: A Case Report
Manish Kiran1, James Kho, Elizabeth Hepburn
1Department of Trauma and Orthopaedics, University Hospitals Bristol and Weston NHS Foundation Trust, Bristol, England.
Case:
We present a case of late onset, bearing wear-induced failure of locking mechanism, resulting in bearing dislocation and rapidly progressive severe metallosis in a medial fixed-bearing (FB) unicompartmental knee arthroplasty (UKA).
Conclusion:
Bearing dislocation is a rare complication of chronic wear in FB UKA, and early recognition is essential to prevent metallic component articulation-induced metallosis and catastrophic failure.