Curtain Flow Column: Optimization of Efficiency and Sensitivity
Published on: June 12, 2016
A systematic investigation of the effect of sample solvent on peak shape in nano- and microflow hydrophilic
Haibin Li1, Chusheng Liu2, Li Zhao3
1Institute of Pharmaceutical Analysis, College of Pharmacy, Jinan University, Guangzhou 510632, China; International Cooperative Laboratory of Traditional Chinese Medicine Modernization and Innovative Drug Development of Chinese Ministry of Education (MOE), College of Pharmacy, Jinan University, Guangzhou 510632, China; Department for Pharmaceutical and Pharmacological Sciences, KU Leuven, Pharmaceutical Analysis, Herestraat 49, Leuven 3000, Belgium.
Abstract:
A frequently encountered problem in the practical application of nano- and microflow hydrophilic interaction chromatography (HILIC) columns is the distortion of peak shapes arising from a mismatch between the sample solvent and the mobile phase. An unmatched or improperly matched sample solvent can distort the peak shape of analytes and influence their retention times, thereby affecting the quality of the resulting chromatogram. In this work, the effect of sample solvent composition (mixtures of acetonitrile, water, methanol and isopropanol in different ratios) and injection volume (20-100 nL) was systematically investigated using a selection of neutral and charged compounds on a series of zwitterionic and charged small I.D. (0.1-0.3 mm) HILIC columns. For retained compounds, pure ACN was demonstrated to be the best sample solvent to obtain narrow peaks, while for compounds that eluted very close to the solvent peak, the peak shape was distorted when the sample solvent consisted of pure ACN. A highly aqueous sample solvent, which interferes with the partitioning of polar analytes into the stationary phase, was demonstrated to be detrimental for the peak shape of retained neutral compounds, while for unretained compounds that do not or hardly interact with the stationary phase, a high amount of water in the sample solvent was not problematic. For charged compounds, water in the sample solvent favored the electrostatic attraction with the stationary phase. Therefore, the retention time of charged analytes was shown to increase with increasing water content in the sample solvent. Even when a large amount of water was present in the sample solvent, the peak shapes of these compounds were still acceptable. For highly polar compounds with a limited solubility in aqueous sample solvents, it was found that a mixture of ACN and MeOH or IPA is a good alternative.
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Principles Of Column Chromatography