Validation of Point-of-Care Ultrasound to Measure Perioperative Edema in Infants With Congenital Heart Disease

Jessica N Persson1,2, Jacqueline Holstein3, Lori Silveira1

  • 1Department of Pediatrics, University of Colorado Anschutz, Aurora, CO, United States.

Frontiers in Pediatrics
|September 9, 2021
PubMed

Insights

Point-of-care ultrasound reliably measures skin edema in infants after cardiac surgery. This method can objectively assess fluid overload, aiding clinical management in neonates and infants.

Area of Science:

  • Pediatric Cardiology
  • Critical Care Medicine
  • Medical Imaging

Background:

  • Fluid overload is a significant post-operative complication in pediatric cardiac surgery, linked to higher morbidity and mortality.
  • Current methods for assessing fluid status lack a gold standard, necessitating objective evaluation tools.

Purpose of the Study:

  • To validate point-of-care ultrasound for measuring skin edema in infants post-cardiac surgery.
  • To evaluate the intra- and inter-user variability of this ultrasound technique.

Main Methods:

  • Prospective cohort study including neonates and infants undergoing cardiac surgery, plus neonatal controls.
  • Skin ultrasound measurements of subcutaneous tissue depth at four body sites, from baseline through post-operative day 3.
  • Intraclass correlation coefficient (ICC) used to assess measurement variability.

Main Results:

  • Skin edema, indicated by increased subcutaneous tissue, was observed starting post-operative day 1 in neonates, persisting through day 3.
  • A similar, though less pronounced, pattern was seen in infants, with near resolution by post-operative day 3.
  • Excellent intra-user (ICC 0.95) and very good inter-user (ICC 0.82) variability confirmed the method's reproducibility.

Conclusions:

  • Point-of-care skin ultrasound is a reproducible and reliable method for assessing subcutaneous tissue and detecting skin edema in infants.
  • This technique offers an objective approach to quantifying extravascular fluid overload in pediatric cardiac surgery patients.
  • Further studies are warranted to correlate ultrasound-detected fluid overload with clinical outcomes.