Related Experiment Video
Updated: Oct 21, 2025

Manufacture of a Multi-Purpose Low-Cost Animal Bench-Model for Teaching Tracheostomy
Published on: May 18, 2019
Bench testing of tracheostomy tube-related insults using an instrumented manikin
Guillaume Cathelain1, Antoine Perrier2,3, Ludivine Folliot4
1CHArt Laboratory, Ecole Pratique Des Hautes Etudes, PSL University, Paris, France.
Purpose:
Choosing the right tracheal tube for the right patient is a daily preoccupation for intensivists and emergency physicians. Tracheal tubes can generate severe complications, which are chiefly due to the pressures applied by the tube to the trachea. We designed a bench study to assess the frequency of pressure levels likely to cause tracheal injury.
Methods:
We tested the pressure applied on the trachea by 17 tube models of a given size range. To this end, we added a pressure sensor to the posterior tracheal wall of a standardized manikin.
Results:
Only 2 of the 17 tubes generated pressures under the threshold likely to induce tracheal injury (30 mmHg/3.99 kPa). The force exerted on the posterior wall of the trachea varied widely across tube models.
Conclusion:
Most models of tracheal tubes resulted in forces applied to the trachea that are usually considered capable of causing tracheal tissue injury.
Level Of Evidence:
Oxford Centre for Evidence-Based Medicine 2011 Levels of Evidence: How common is the problem?: step 1; Is this diagnostic or monitoring test accurate? (Diagnosis) step 5; What will happen if we do not add a therapy? (Prognosis) n/a; Does this intervention help? (Treatment Benefits) step 5; What are the COMMON harms?(Treatment Harms) step 5; What are the RARE harms? (Treatment Harms) step 5; Is this (early detection) test worthwhile? (Screening) step 5.
Related Concept Videos
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes: